These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Point-of-care β-hydroxybutyrate measurement for the diagnosis of feline diabetic ketoacidaemia. Author: Zeugswetter FK, Rebuzzi L. Journal: J Small Anim Pract; 2012 Jun; 53(6):328-31. PubMed ID: 22533366. Abstract: OBJECTIVES: To evaluate accuracy and precision of a hand-held ketone meter measuring β-hydroxybutyrate and to determine its diagnostic performance to rule out ketoacidaemia in diabetic cats. METHODS: The ketone meter was validated by calculating within-day precision at different β-hydroxybutyrate concentrations and by comparison with a laboratory method. To determine its diagnostic performance to diagnose ketoacidaemia, 217 sets of data (venous blood gas analysis and β-hydroxybutyrate measurements) were retrospectively analysed. Sensitivities and specificities were calculated with the help of receiver-operating characteristic curves. RESULTS: The ketone meter reliably detected β-hydroxybutyrate at concentrations >0·1 mmol/L and reproducibility was acceptable. Measurements highly correlated with laboratory results (r=0·97; P<0·001), but a significant negative bias was found at high concentrations. A β-hydroxybutyrate concentration of >2·55 mmol/L had a sensitivity of 94% and a specificity of 68% for diagnosing ketoacidaemia. Many cats with high β-hydroxybutyrate concentrations and normal blood pH had an elevated chloride gap suggestive of superimposed hypochloraemic metabolic alkalosis. CLINICAL SIGNIFICANCE: The commercially available point-of-care ketone meter Precision Xtra is a valid tool to measure β-hydroxybutyrate in diabetic cats. Concentration <2·55 mmol/L enable ketoacidaemia to be excluded and should lead to redirection of differential diagnoses.[Abstract] [Full Text] [Related] [New Search]