These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Atraumatic measurement of intracranial pressure]. Author: Russegger L, Ennemoser O. Journal: Wien Klin Wochenschr; 1990 Sep 28; 102(18):543-7. PubMed ID: 2264345. Abstract: This paper presents a new method for measuring intracranial pressure (ICP), which is completely non-invasive by using an isolated "headband electrode". The head itself - as a dielectric--forms a capacitor. A high frequency alternating voltage field is applied to this system (1-1.5 MHz) whereby a measuring signal is obtained which depends on the dielectric properties of the head. These, in turn, primarily depend on the filling status of the intracranial space. In addition, a pulse wave can be plotted as a result of cardiac and respiratory cycles. This combined information permits the calculation of actual ICP values with the help of a simple Basic PC programme. The measuring system was tested in 20 patients with intracranial space-occupying lesions. Their ICP values were known from invasive measurement and the obtained correlation was R = 0.983.[Abstract] [Full Text] [Related] [New Search]