These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Development and validation of a simple and fast HPLC method for determination of lovastatin, pravastatin and simvastatin.
    Author: Silva TD, Oliveira MA, de Oliveira RB, Vianna-Soares CD.
    Journal: J Chromatogr Sci; 2012 Oct; 50(9):831-8. PubMed ID: 22689902.
    Abstract:
    Statins are effective and often-prescribed drugs for the treatment of hypercholesterolemia. This study shows a simple and fast method validation by reversed-phase high-performance liquid chromatography in the linear range 28 to 52 µg/mL to quantify lovastatin, pravastatin sodium or simvastatin in bulk drug or dosage forms. Statins were determined using a C8 endcapped column (250 × 4 mm, 5 µm), isocratic mobile phase of acetonitrile and 0.1% phosphoric acid (65:35), 30°C, ultraviolet-diode array detection at λ 238 nm and 1.5 mL/min flow for lovastatin and simvastatin and 1.0 mL/min for pravastatin sodium. The developed method is fast, simple, reliable and shows appropriate linearity (r > 0.999), accuracy (98.8-101.6%), precision (relative standard deviation <2%) and selectivity toward placebo and/or degradation products in very similar chromatographic conditions for all statins.
    [Abstract] [Full Text] [Related] [New Search]