Title: Development and validation of a simple and fast HPLC method for determination of lovastatin, pravastatin and simvastatin. Author: Silva TD, Oliveira MA, de Oliveira RB, Vianna-Soares CD. Journal: J Chromatogr Sci; 2012 Oct; 50(9):831-8. PubMed ID: 22689902. Abstract: Statins are effective and often-prescribed drugs for the treatment of hypercholesterolemia. This study shows a simple and fast method validation by reversed-phase high-performance liquid chromatography in the linear range 28 to 52 µg/mL to quantify lovastatin, pravastatin sodium or simvastatin in bulk drug or dosage forms. Statins were determined using a C8 endcapped column (250 × 4 mm, 5 µm), isocratic mobile phase of acetonitrile and 0.1% phosphoric acid (65:35), 30°C, ultraviolet-diode array detection at λ 238 nm and 1.5 mL/min flow for lovastatin and simvastatin and 1.0 mL/min for pravastatin sodium. The developed method is fast, simple, reliable and shows appropriate linearity (r > 0.999), accuracy (98.8-101.6%), precision (relative standard deviation <2%) and selectivity toward placebo and/or degradation products in very similar chromatographic conditions for all statins.[Abstract] [Full Text] [Related] [New Search]