These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Detection of gliadin in foods using a quartz crystal microbalance biosensor that incorporates gold nanoparticles.
    Author: Chu PT, Lin CS, Chen WJ, Chen CF, Wen HW.
    Journal: J Agric Food Chem; 2012 Jul 04; 60(26):6483-92. PubMed ID: 22694361.
    Abstract:
    This work develops a label-free gliadin immunosensor that is based on changes in the frequency of a quartz crystal microbalance (QCM) chip. A higher sensitivity was obtained by applying 25 nm gold nanoparticles (AuNPs) to the surface of a bare QCM electrode. Subsequently, chicken anti-gliadin antibodies (IgY) were immobilized directly on the AuNP-modified surface by cross-linking amine groups in IgY with glutaraldehyde. Experimental results revealed that the change in frequency exhibited when 2 ppm gliadin was bound to the AuNP-modified electrode was 35 Hz (48%) greater than that of the bare gold electrode. The linear dynamic range in 60% ethanol was from 1 × 10(1) to 2 × 10(5) ppb gliadin, and the calculated limit of detection (LOD) was 8 ppb. The entire detection process was completed in 40 min and was highly repeatable. Additionally, the AuNP-modified QCM system generated results in the detection of gliadin in 10 commercial food products that were consistent with those obtained using an AOAC-approved gliadin kit. In conclusion, the QCM platform provides a potential alternative means of ensuring that people with wheat allergies and celiac patients have access to gliadin-free food.
    [Abstract] [Full Text] [Related] [New Search]