These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A new cable-tie based sternal closure system: description of the device, technique of implantation and first clinical evaluation. Author: Grapow MT, Melly LF, Eckstein FS, Reuthebuch OT. Journal: J Cardiothorac Surg; 2012 Jun 25; 7():59. PubMed ID: 22731778. Abstract: BACKGROUND: Wire closure still remains the preferred technique despite reasonable disadvantages. Associated complications, such as infection and sternal instability, cause time- and cost-consuming therapies. We present a new tool for sternal closure with its first clinical experience and results. METHODS: The sternal ZipFix(TM) System is based on the cable-tie principle. It primarily consists of biocompatible Poly-Ether-Ether-Ketone implants and is predominantly used peristernally through the intercostal space. The system provides a large implant-to-bone contact for better force distribution and for avoiding bone cut through. RESULTS: 50 patients were closed with the ZipFix(TM) system. No sternal instability was observed at 30 days. Two patients developed a mediastinitis that necessitated the removal of the device; however, the ZipFix(TM) were intact and the sternum remained stable. CONCLUSIONS: In our initial evaluation, the short-term results have shown that the sternal ZipFix(TM) can be used safely and effectively. It is fast, easy to use and serves as a potential alternative for traditional wire closure.[Abstract] [Full Text] [Related] [New Search]