Title: Investigation of alpha and beta self-absorption factors in the calibration of water sample measurements. Author: Valentin CP, Kratky J, Brandl A. Journal: Health Phys; 2012 Aug; 103(2 Suppl 2):S124-30. PubMed ID: 22739965. Abstract: Detector efficiencies and self-absorption factors for a variety of sample preparation methods and naturally occurring dissolved ions have been studied for the gross alpha and gross beta measurement of water samples. The most consistent results are obtained for samples prepared by evaporation of small volumes of water under an infrared lamp. Due to its uniform and homogeneous deposition in the cupped stainless steel planchet, CaCO₃ was found to best serve as the dry residue radionuclide carrier in this study. Detector efficiencies and self-absorption factors for gross alpha and gross beta measurements closely follow the expected exponential and logarithmic dependence, respectively.[Abstract] [Full Text] [Related] [New Search]