These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Validated RP-HPLC/UV method for the quantitation of abiraterone in rat plasma and its application to a pharmacokinetic study in rats. Author: Kumar SV, Rudresha G, Gurav S, Zainuddin M, Dewang P, Kethiri RR, Rajagopal S, Mullangi R. Journal: Biomed Chromatogr; 2013 Feb; 27(2):203-7. PubMed ID: 22763809. Abstract: A novel, simple, specific, sensitive and reproducible high-performance liquid chromatography (HPLC) assay method has been developed and validated for the estimation of abiraterone (ART) in rat plasma. The analytical procedure involves extraction of ART and diclofenac (internal standard, IS) from rat plasma with a simple liquid-liquid extraction process. The chromatographic analysis was performed on a Waters Alliance system with a Betasil C(18) column maintained at ambient room temperature and an isocratic mobile phase [acetonitrile-water-10 mm potassium dihydrogen phosphate (pH 3.0), 55:5:40, v/v/v] at a flow rate of 1.00 mL/min with a total run time of 10 min. The eluate was monitored using an UV detector set at 255 nm. Method validation was performed as per FDA guidelines and the results met the acceptance criteria. The calibration curve was linear over a concentration range of 93.4-3251 ng/mL (r(2) = 0.997). The intra- and inter-day precisions were 0.56-4.98 and 3.03-7.18, respectively, in rat plasma. The validated HPLC method was successfully applied to a pharmacokinetic study of ART in rats.[Abstract] [Full Text] [Related] [New Search]