These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Spinal canal surrogate for testing intradural implants.
    Author: Oya H, Howard MA, Shurig R, Gillies GT.
    Journal: J Med Eng Technol; 2012 Nov; 36(8):407-10. PubMed ID: 22953718.
    Abstract:
    We have designed, built and tested an anthropomorphic-scale surrogate spinal canal, for use in preliminary evaluations of the performance characteristics of a novel intradural spinal cord stimulator. The surrogate employs a silicone mock spinal cord with semi-major and semi-minor diameters of 10 and 6 mm, respectively, commensurate with those of actual thoracic-level spinal cord. The axial restoring force provided by the 300 µm thick silicone denticulate ligament constructs on the mock cord is ~ 0.32 N mm(-1) over a 1.5 mm range of displacement, which is within a factor of 2 of that measured by others in human cadaver specimens. Examples of testing protocols of prototype intradural stimulators that employ this device are discussed.
    [Abstract] [Full Text] [Related] [New Search]