These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Equivalency of protein determination in feed on Bran & Luebbe Traacs 800 autoanalyzer with AOAC semiautomated method: minicollaborative study. Author: Kane PF, Orcutt A. Journal: J Assoc Off Anal Chem; 1990; 73(1):31-4. PubMed ID: 2312510. Abstract: A new generation of autoanalyzers was collaboratively studied for equivalency with instrumentation described in the AOAC method for crude protein in animal feeds. Twenty predigested samples were analyzed by 4 collaborators using standards that were provided. Means were comparable; the overall average difference was 0.07% protein. Analysis of variance indicated no evidence for an instrument difference. The method describing use of the Traacs 800 autoanalyzer as an alternative system has been adopted official first action.[Abstract] [Full Text] [Related] [New Search]