These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Pt/graphene-CNTs nanocomposite based electrochemical sensors for the determination of endocrine disruptor bisphenol A in thermal printing papers.
    Author: Zheng Z, Du Y, Wang Z, Feng Q, Wang C.
    Journal: Analyst; 2013 Jan 21; 138(2):693-701. PubMed ID: 23187892.
    Abstract:
    A facile and green method was developed to synthesize the graphene-carbon nanotubes (Gr-CNTs) nanocomposite with a sandwich lamination structure. Pt nanoparticles were loaded on the as-synthesized Gr-CNTs nanocomposite to prepare an electrochemical sensor for determining bisphenol A (BPA) in thermal printing paper. The electrochemical behavior of BPA on the Pt/Gr-CNTs nanocomposite was investigated by cyclic voltammetry (CV) and chronocoulometry (CC). The direct determination of BPA was accomplished by using differential pulse voltammetry (DPV) under optimized conditions. The oxidation peak current was proportional to the BPA concentration in the range from 6.0 × 10(-8) to 1.0 × 10(-5) M and 1.0 × 10(-5) to 8.0 × 10(-5) M with a correlation coefficient of 0.987 and 0.998, respectively. The detection limit was 4.2 × 10(-8) M (S/N = 3). The fabricated electrode showed good reproducibility, stability and selectivity. The proposed method was successfully applied to determine BPA in thermal printing papers samples and the results were satisfactory.
    [Abstract] [Full Text] [Related] [New Search]