Search MEDLINE/PubMed
Title: Comparison of mycophenolic acid concentrations determined by a new PETINIA assay on the dimension EXL analyzer and a HPLC-UV method. Author: Dasgupta A, Tso G, Chow L. Journal: Clin Biochem; 2013 May; 46(7-8):685-7. PubMed ID: 23219740. Abstract: OBJECTIVE: Mycophenolic acid requires routine therapeutic drug monitoring. We evaluated the suitability of a new PETINIA (particle enhanced turbidimetric inhibition immunoassay) assay on the Dimension EXL analyzer for monitoring of mycophenolic acid by comparing values obtained by this assay with a HPLC-UV method. DESIGN AND METHODS: Mycophenolic acid concentrations determined in sera of 60 organ transplant recipients using high performance liquid chromatography combined with ultraviolet detection (HPLC-UV, reference method) and the new immunoassay on the Dimension RxL analyzer. RESULTS: The within and between run precision of the new PETINIA assay was <10%. The assay was linear for a mycophenolic acid concentration up to 30μg/mL. When mycophenolic acid concentrations in 60 transplant recipients obtained by the HPLC-UV (x-axis) method were compared with corresponding values obtained by the PETINIA assay (y-axis), the following regression equation was obtained: y=1.1204 x+0.0881 (r=0.983, n=60). CONCLUSIONS: If PETINIA assay is used for therapeutic drug monitoring of mycophenolic acid, caution must be exercised in interpreting serum mycophenolic acid level due to observed positive bias.[Abstract] [Full Text] [Related] [New Search]