These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Validity assessment of a portable bioimpedance scale to estimate body fat percentage in white and African-American children and adolescents. Author: Barreira TV, Staiano AE, Katzmarzyk PT. Journal: Pediatr Obes; 2013 Apr; 8(2):e29-32. PubMed ID: 23239610. Abstract: OBJECTIVE: The objective of the study was to determine accuracy of the Tanita SC-240 body composition analyser to measure paediatric percent body fat (%BF). METHODS: Eighty-nine African-American and white 5-18-year-olds participated in this study. %BF was estimated by dual-energy X-ray absorptiometry (DXA) and by the Tanita SC-240. RESULTS: Overall %BF was 33.5 ± 10.5% (Tanita SC-240) vs. 34.5 ± 8.7% (DXA). There was no significant difference between the two measures (P = 0.52, average error = -1.0%, average absolute error = 3.9%). The Tanita mean %BF estimates significantly differed from the DXA mean %BF in white boys (P = 0.001, Cohen's d = 0.40) and white girls (P = 0.006, Cohen's d = 0.48), but differences were of small effect. No differences in %BF estimates were found for African-American boys or girls. CONCLUSIONS: In this sample, the Tanita SC-240 demonstrated acceptable accuracy for estimating %BF when compared with DXA, supporting its use in field studies.[Abstract] [Full Text] [Related] [New Search]