  • Title: Preventing lower cranial nerve injuries during fourth ventricle tumor resection by utilizing intraoperative neurophysiological monitoring.
    Author: Jahangiri FR, Minhas M, Jane J.
    Journal: Neurodiagn J; 2012 Dec; 52(4):320-32. PubMed ID: 23301282.
    Abstract:
    We present two cases illustrating the benefit of utilizing intraoperative neurophysiological monitoring (IONM) for prevention of injuries to the lower cranial nerves during fourth ventricle tumor resection surgeries. Multiple cranial nerve nuclei are located on the floor of the fourth ventricle with a high risk of permanent damage. Two male patients (ages 8 and 10 years) presented to the emergency department and had brain magnetic resonance imaging (MRI) scans showing brainstem/fourth ventricle tumors. During surgery, bilateral posterior tibial and median nerve somatosensory evoked potentials (SSEPs); four-limb and cranial nerves transcranial electrical motor evoked potentials (TCeMEPs); brainstem auditory evoked responses (BAERs); and spontaneous electromyography (s-EMG) were recorded. Electromyography (EMG) was monitored bilaterally from cranial nerves V VII, IX, X, XI, and XII. Total intravenous anesthesia was used. Neuromuscular blockade was used only for initial intubation. Pre-incision baselines were obtained with good morphology of waveforms. After exposure the floor of the fourth ventricle was mapped by triggered-EMG (t-EMG) using 0.4 to 1.0 mA. In both patients the tumor was entangled with cranial nerves VII to XII on the floor of the fourth ventricle. The surgeon made the decision not to resect the tumor in one case and limited the resection to 70% of the tumor in the second case on the basis of neurophysiological monitoring. This decision was made to minimize any post-operative neurological deficits due to surgical manipulation of the tumor involving the lower cranial nerves. Intraoperative spontaneous and triggered EMG was effectively utilized in preventing injuries to cranial nerves during surgical procedures. All signals remained stable during the surgical procedure. Postoperatively both patients were well with no additional cranial nerve weakness. At three months follow-up, the patients continued to have no deficits.
    [Abstract] [Full Text] [Related] [New Search]