These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Intrapartum heart rate ambiguity: a comparison of cardiotocogram and abdominal fetal electrocardiogram with maternal electrocardiogram.
    Author: Reinhard J, Hayes-Gill BR, Schiermeier S, Hatzmann H, Heinrich TM, Louwen F.
    Journal: Gynecol Obstet Invest; 2013; 75(2):101-8. PubMed ID: 23328351.
    Abstract:
    OBJECTIVE/AIMS: To investigate the presence of signal ambiguity of intrapartum fetal heart rate (FHR) monitoring during delivery by comparing simultaneous cardiotocogram (CTG), abdominal fetal electrocardiogram (ECG) with continuous maternal ECG. METHODS: A total of 144 simultaneous CTG (Corometrics 250 series), abdominal fetal ECG (Monica -AN24™) and maternal ECG (Monica AN24™) recordings were evaluated. MAIN OUTCOME MEASURES: When the FHR is within 5 bpm of the maternal heart rate (MHR) acquired from the ECG, it is classified as 'MHR/FHR ambiguity'. Statistical analyses were performed with Fisher's exact test and the Wilcoxon signed-rank test. RESULTS: Comparison of abdominal fetal ECG against CTG demonstrates significantly less 'MHR/FHR ambiguity' in both the first stage (mean 0.70 vs. 1.22%, p < 0.001) and second stage of labour (mean 3.30 vs. 6.20%, p < 0.001). CONCLUSION: Intrapartum FHR monitoring in daily practice via the CTG modality provides significantly more 'MHR/FHR ambiguity' than abdominal fetal ECG, which also provides additional information on the MHR.
    [Abstract] [Full Text] [Related] [New Search]