These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Low-power system-on-chip implementation for respiratory rate detection and transmission. Author: Padasdao B, Yee R, Boric-Lubecke O. Journal: Annu Int Conf IEEE Eng Med Biol Soc; 2012; 2012():4501-4. PubMed ID: 23366928. Abstract: Recent biosensors can measure respiratory rate non-invasively, but limits patient mobility or requires regular battery replacement. Respiratory effort, which can scavenge mW, may power the sensor, but requires minimal sensor power usage. This paper demonstrates feasibility of respiratory rate measurement by using a comparator instead of ADC. A low-power system-on-chip can implement respiratory rate detection and wireless data transmission with a total power consumption under 82 µW. This approach produces significant power savings, and transmission uses under 30% of total power consumption.[Abstract] [Full Text] [Related] [New Search]