Title: Sequential algorithm for the detection of the shockable rhythms in electrocardiogram. Author: Jeong JW, Lee IB, Song Y, Jang Y, Noh HW, Lee S. Journal: Annu Int Conf IEEE Eng Med Biol Soc; 2012; 2012():5082-5. PubMed ID: 23367071. Abstract: We suggest a sequential algorithm for the detection of the ventricular fibrillation (VF) and ventricular tachycardia (VT) of a rate above 180 bpm, so called shockable rhythms. The built-in algorithm for ECG analysis embedded in the portable bio-signal sensing module is aimed to discriminate between shockable and non-shockable rhythms and its accuracy is analyzed. An algorithm for VF/VT detection is proposed to analyze every 1 s ECG episode using the past 8 s episodes. The method is tested with 844,587 ECG episodes from the widely accepted databases. A sensitivity of 86.8 % and a specificity of 99.4 % were obtained and compared with the previous results.[Abstract] [Full Text] [Related] [New Search]