These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Selective left ventricular sensing lead implantation to overcome undersensing of ventricular fibrillation during implantable cardioverter defibrillator implantation. Author: Steinberg C, Philippon F, O'Hara G, Molin F. Journal: Can J Cardiol; 2013 Jun; 29(6):751.e3-6. PubMed ID: 23395280. Abstract: Accurate sensing of malignant arrhythmia is critical for the appropriate delivery of therapy from implantable cardioverter defibrillators, and undersensing of ventricular tachyarrhythmias can have catastrophic consequences. Here, we present an unusual case of ventricular fibrillation undersensing from the right ventricular lead at multiple different implantation sites because of very low amplitude voltage signals during induced ventricular fibrillation. A left ventricular sensing electrode was implanted to allow correct sensing and therapy delivery.[Abstract] [Full Text] [Related] [New Search]