Title: Improved slice-selective adiabatic excitation. Author: Balchandani P, Glover G, Pauly J, Spielman D. Journal: Magn Reson Med; 2014 Jan; 71(1):75-82. PubMed ID: 23401184. Abstract: PURPOSE: The purpose of this work is to design an improved Slice-selective Tunable-flip AdiaBatic Low peak-power Excitation (STABLE) pulse with shorter duration and increased off-resonance immunity to make it suitable for use in a greater range of applications and at higher field strengths. An additional aim is to design a variant of this pulse to achieve B1 -insensitive, fat-suppressed excitation. METHODS: The adiabatic SLR algorithm was used to generate a more uniform spectral pulse envelope for this improved radiofrequency pulse for adiabatic slice-selective excitation, called STABLE-2. Pulse parameters were adjusted to design a version of STABLE-2 with a spectral null centered on lipids. RESULTS: In vivo images obtained of the human brain at 3 and 7 T demonstrate that STABLE-2 provides robust, uniform, slice-selective excitation over a range of B1 values. Phantom and in vivo knee images obtained at 3 T demonstrate the effectiveness of STABLE-2 for fat suppression. CONCLUSIONS: STABLE-2 achieves B1 -insensitive slice-selective excitation while providing greater off-resonance immunity and a shorter pulse duration, when compared to the original STABLE pulse. In particular, the 9.8-ms STABLE-2 pulse provides slice selectivity over 120 Hz whereas the 21-ms STABLE pulse is limited to 80 Hz off-resonance. B1 -Insensitive fat-suppressed excitation may also be achieved by using a variant of this pulse.[Abstract] [Full Text] [Related] [New Search]