These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Evaluation of buprenorphine LUCIO immunoassay versus GC-MS using urines from a workplace drug testing program.
    Author: Alves MN, Piccinotti A, Tameni S, Polettini A.
    Journal: J Anal Toxicol; 2013 Apr; 37(3):175-8. PubMed ID: 23408251.
    Abstract:
    The buprenorphine (BUP) LUCIO Nal Von Minden screening assay was evaluated. Urine samples from subjects enrolled in a workplace drug testing program were screened according to the manufacturer's instruction using a Roche COBAS Integra 800 analyser. For gas chromatography-mass spectrometry (GC-MS) confirmatory analysis, samples were submitted to enzymatic hydrolysis with β-glucuronidase and mixed-mode solid-phase extraction. Imprecision (coefficient of variation) for 3.0, 7.0, and 13.0 ng/mL calibrators varied within 2.8-8.7 intra-day (n = 20) and 7.7-8.6 inter-day (n = 19). Inaccuracy (bias) was between -5.6-30.5 intra-day and -13.2-4.2 inter-day. At the 5 ng/mL cut-off, the immunoassay showed 100% sensitivity and 88% specificity, with an overall agreement of 94% between immunoassay and GC-MS. Raising the cut-off to 10 ng/mL provided an identical overall agreement between immunoassay and GC-MS (94%), despite the decrease in sensitivity (90%) and the increase in specificity (100%). According to these results, the BUP LUCIO Nal Von Minden screening assay provides adequate sensitivity and specificity for BUP screening in urine samples using a cut-off concentration of 5 ng/mL.
    [Abstract] [Full Text] [Related] [New Search]