These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: The difference in intraocular pressure readings between 3 applanation tonometers is independent of central corneal thickness, in glaucomatous and nonglaucomatous eyes. Author: Baily C, Dooley I, Collins N, Hickey-Dwyer M. Journal: J Glaucoma; 2014 Dec; 23(9):620-3. PubMed ID: 23429625. Abstract: PURPOSE: To look at tonometry differences between 3 tonometers and to correlate this difference with central corneal thickness (CCT) in glaucomatous and nonglaucomatous eyes. METHODS: A total of 145 patients were recruited (39 with glaucoma). Intraocular pressure (IOP) was carried out using the Tonosafe (TS), Tono-Pen XL (TP), and Goldmann applanation tonometer (GAT). CCT was also performed using a digital pachymeter. RESULTS: The TS underestimated IOP when compared with the GAT by a mean (±SD) of 1.03±2.14 mm Hg (P<0.001). The TP when compared to the GAT overestimated IOP by a mean (±SD) of +1.09±6.38 mm Hg (P=0.071). We found only 31% of TP readings to be within ±2 mm Hg of the GAT readings and only 60% to be within ±4 mm Hg of the GAT readings. Correlation was poor (F=0.247) between the GAT and TP with the coefficient of determination being 0.061. The difference in IOP was independent of CCT and a diagnosis of glaucoma. CONCLUSIONS: The TS, when compared with the GAT, underestimates IOP with a tendency for this to increase in the higher range of IOP readings. The TP is poorly correlated with the GAT.[Abstract] [Full Text] [Related] [New Search]