These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: HPLC-NMR revisited: using time-slice high-performance liquid chromatography-solid-phase extraction-nuclear magnetic resonance with database-assisted dereplication. Author: Johansen KT, Wubshet SG, Nyberg NT. Journal: Anal Chem; 2013 Mar 19; 85(6):3183-9. PubMed ID: 23432092. Abstract: Time-based trapping of chromatographically separated compounds onto solid-phase extraction (SPE) cartridges and subsequent elution to NMR tubes was done to emulate the function of HPLC-NMR for dereplication purposes. Sufficient mass sensitivity was obtained by use of a state-of-the-art HPLC-SPE-NMR system with a cryogenically cooled probe head, designed for 1.7 mm NMR tubes. The resulting (1)H NMR spectra (600 MHz) were evaluated against a database of previously acquired and prepared spectra. The in-house-developed matching algorithm, based on partitioning of the spectra and allowing for changes in the chemical shifts, is described. Two mixtures of natural products were used to test the approach: an extract of Carthamus oxyacantha (wild safflower), containing an array of spiro compounds, and an extract of the endophytic fungus Penicillum namyslowski, containing griseofulvin and analogues. The database matching of the resulting spectra positively identified expected compounds, while the number of false positives was few and easily recognized.[Abstract] [Full Text] [Related] [New Search]