  • Title: Response of six neutron survey meters in mixed fields of fast and thermal neutrons.
    Author: Kim SI, Kim BH, Chang I, Lee JI, Kim JL, Pradhan AS.
    Journal: Radiat Prot Dosimetry; 2013 Oct; 156(4):518-24. PubMed ID: 23620566.
    Abstract:
    Calibration neutron fields have been developed at KAERI (Korea Atomic Energy Research Institute) to study the responses of commonly used neutron survey meters in the presence of fast neutrons of energy around 10 MeV. The neutron fields were produced by using neutrons from the (241)Am-Be sources held in a graphite pile and a DT neutron generator. The spectral details and the ambient dose equivalent rates of the calibration fields were established, and the responses of six neutron survey meters were evaluated. Four single-moderator-based survey meters exhibited an under-responses ranging from ∼9 to 55 %. DINEUTRUN, commonly used in fields around nuclear reactors, exhibited an over-response by a factor of three in the thermal neutron field and an under-response of ∼85 % in the mixed fields. REM-500 (tissue-equivalent proportional counter) exhibited a response close to 1.0 in the fast neutron fields and an under-response of ∼50 % in the thermal neutron field.
    [Abstract] [Full Text] [Related] [New Search]