These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Molecularly imprinted polyaniline-polyvinyl sulphonic acid composite based sensor for para-nitrophenol detection. Author: Roy AC, Nisha VS, Dhand C, Ali MA, Malhotra BD. Journal: Anal Chim Acta; 2013 May 13; 777():63-71. PubMed ID: 23622966. Abstract: We report results of the studies relating to the fabrication and characterization of a conducting polymer based molecularly imprinted para-nitrophenol (PNP) sensor. A water pollutant, para-nitrophenol is electrochemically imprinted with polyvinyl sulphonic acid (PVSA) doped polyaniline onto indium tin oxide (ITO) glass substrate. This PNP imprinted electrode (PNPI-PANI-PVSA/ITO) prepared via chronopotentiometric polymerization and over-oxidation is characterized by Fourier transform infra-red spectroscopy (FT-IR), UV-visible (UV-vis) spectroscopy, contact angle (CA), scanning electron microscopy (SEM), cyclic voltammetry (CV) and differential pulse voltammetry (DPV) studies. The response studies of PNPI-PANI-PVSA/ITO electrode carried out using DPV reveal a lower detection limit of 1×10(-3) mM, improved sensitivity as 1.5×10(-3) A mM(-1) and stability of 45 days. The PNPI-PANI-PVSA/ITO electrode shows good precision with relative standard deviation of 2.1% and good reproducibility with standard deviation of 3.78%.[Abstract] [Full Text] [Related] [New Search]