These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: The effect of increased centrifugation temperature on the quality of red-blood-cell concentrates of automated whole blood processing.
    Author: Weinigel C, Rummler S, Barz D.
    Journal: Vox Sang; 2013 Oct; 105(3):205-9. PubMed ID: 23663207.
    Abstract:
    BACKGROUND AND OBJECTIVES: There are manual and automated methods to separate whole blood (WB) available. The Atreus whole blood processing system is an automated method, which combines centrifugation and expression of components into a single device. A major difference to conventional methods is that centrifugation temperature is not controlled at 22°C. The aim of this study was to examine the influence of increased centrifugation temperatures on the quality of red-blood-cell concentrates (RCC) after active cooling of WB prior to processing. MATERIALS AND METHODS: A total of 28 WB were processed: 16 at centrifugation temperatures of up to 28°C (1st protocol) and 12 at 34°C (2nd protocol). RCC quality parameters were tested weekly for 42 days. RESULTS: Red-blood-cell concentrates (RCC) quality complied with the European and German guidelines. Haemolysis was not significantly different throughout storage. Significant statistical differences were detected between both protocols in potassium concentration at the end of storage and in ATP levels at the day of processing. CONCLUSION: Centrifugation temperatures of up to 34°C are well tolerated by the red blood cells with minimal interference with the RCC quality parameters.
    [Abstract] [Full Text] [Related] [New Search]