These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Impact of mass spectrometry by MALDI-TOF MS for the rapid identification of aerobic and anaerobic bacteria of clinical importance].
    Author: Legarraga P, Moraga M, Lam M, Geoffroy E, Zumarán C, García P.
    Journal: Rev Chilena Infectol; 2013 Apr; 30(2):140-6. PubMed ID: 23677152.
    Abstract:
    BACKGROUND: MALDI-TOF MS (Matrix Assisted Laser Desorption Ionization -Time of Flight Mass Spectrometry) technology, recently introduced in the microbiology laboratory has proven to be a precise and rapid method for bacterial identification. OBJECTIVE: To evaluate the performance, costs associated and turnaround time of MALDI-TOF in a routine laboratory. MATERIAL AND METHOD: Five hundred and sixty one clinical isolates (281 aerobes and 280 anaerobes) previously identified by conventional methods were evaluated. Discordances were resolved by means of 16S rRNA sequencing. RESULTS: MALDI-TOF identified 95, 7% of the aerobes isolates and 86, 4% of the anaerobes. The groups with better performance were the enterobacteriacea and Bacteroides spp with 95% and 100% identification at the species level. The error rate of MALDI-TOF and conventional methods compared to sequencing was 0, 39% and 9, 4% respectively. The costs associated were 8 times lower with a turnaround time of 6 hours. CONCLUSION: MALDI-TOF proved to be simple, precise and less expensive technology compared to the traditional methods.
    [Abstract] [Full Text] [Related] [New Search]