  • Title: Comparison of calibrated chromogenic anti-Xa assay and PT tests with LC-MS/MS for the therapeutic monitoring of patients treated with rivaroxaban.
    Author: Douxfils J, Tamigniau A, Chatelain B, Chatelain C, Wallemacq P, Dogné JM, Mullier F.
    Journal: Thromb Haemost; 2013 Oct; 110(4):723-31. PubMed ID: 23846172.
    Abstract:
    Possibilities to monitor rivaroxaban therapy could be useful in certain circumstances. Prothrombin time (PT) or chromogenic anti-Xa assays such as the Biophen Direct Factor Xa Inhibitor® (DiXaI) have been proposed to estimate rivaroxaban concentrations but are mainly based on in vitro studies. The study aim was to compare PT and Biophen DiXaI® measurements with liquid chromatography-tandem mass spectrometry (LC-MS/MS) measurements in plasma samples from patients treated with Xarelto®. Fifty-two plasma samples were included. PT was performed using Innovin® and Triniclot PT Excel S®. Biophen DiXaI® was performed according to instructions from the manufacturer. The rivaroxaban plasma concentration ranged between 0 and 485 ng/ml as measured by LC-MS/MS. The limits of quantification were 30 ng/ml and 5 ng/ml for Biophen DiXaI® and LC-MS/MS, respectively. The linear correlation between Biophen DiXaI® and LC-MS/MS analyses was high for all rivaroxaban concentrations (r² = 0.95). For concentrations ≤100 ng/ml, r²-value was 0.83. The Bland-Altman analysis showed a mean difference of -16 ng/ml (SD: 25 ng/ml). The PT methods did not correlate well with plasma concentrations measured by LC-MS/MS (r² ≈ 0.60). In conclusion, the important inter-individual variability and the poor correlation with LC-MS/MS preclude the use of PT to estimate rivaroxaban concentrations. Thanks to its small inter-individual variability and good agreement with LC-MS/MS measurements, we recommend the use of Biophen DiXaI® assays to estimate concentrations of rivaroxaban >30 ng/ml. Quantification of low rivaroxaban levels (<30 ng/ml) requires the LC-MS/MS method.
    [Abstract] [Full Text] [Related] [New Search]