These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Glucose microfluidic biosensors based on reversible enzyme immobilization on photopatterned stimuli-responsive polymer.
    Author: Xiong M, Gu B, Zhang JD, Xu JJ, Chen HY, Zhong H.
    Journal: Biosens Bioelectron; 2013 Dec 15; 50():229-34. PubMed ID: 23867353.
    Abstract:
    In this paper, we demonstrate a new strategy for replaceable enzymatic microreactor based on a switchable wettability interface of poly(N-isopropylacrylamide) (PNIPAAm). PNIPAAm porous polymer monolith (PPM) with 3D macroporous framework is photopolymerized in glass microchip within 30 s. The PNIPAAm PPM not only shows its reversible swelling/shrinking property at the different temperature around the lower critical solution temperature (LCST), but also shows reversible hydrophilicity/hydrophobicity corresponding to its swelling/shrinking status. Based on these properties, a biocompatible and replaceable on-chip enzymatic microreactor has been successfully built by means of the reversible adsorption and release of glucose oxidase (GOx) on the robust and stable matrix. Coupled with a carbon fiber microelectrode as electrochemical detector, the microreactor has been successfully employed for detection of glucose with a linear range from 0.05 to 5 mM. This approach may provide a promising way for high efficient and renewable microreactors that will find wide application in clinical diagnosis, biochemical synthesis/analysis, and proteomic research.
    [Abstract] [Full Text] [Related] [New Search]