These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Liquid chromatography-tandem mass spectrometry of porphyrins and porphyrinogens in biological materials: separation and identification of interfering poly(ethylene) glycol by travelling wave ion mobility spectrometry/tandem mass spectrometry. Author: Benton CM, Lim CK, Moniz C, Jones DJ. Journal: Biomed Chromatogr; 2013 Dec; 27(12):1782-7. PubMed ID: 23893773. Abstract: Biological and clinical samples for porphyrin and porphyrinogen analyses by liquid chromatography-tandem mass spectrometry (LC-MS/MS) are often contaminated with poly(ethylene)glycol (PEG), which complicates the interpretation of mass spectra and characterisation of new porphyrin metabolites. Two contaminating PEG molecules (m/z 833 and m/z 835) were completely separated from uroporphyrin I (m/z 831) by travelling wave ion mobility spectrometry and characterised by tandem mass spectrometry. One of the PEG species (m/z 835) also co-eluted with uroporphyrinogen I (m/z 837) and was unresolvable by travelling wave ion mobility spectrometry/MS, therefore contaminating the MS/MS mass spectra owing to isotope distribution. These PEG species, with the [M + H](+) ions at m/z at 833 and/or m/z 835, co-eluted with uroporphyrin I and uroporphyrinogen I by LC-MS/MS and could be wrongly identified as uroporphomethenes.[Abstract] [Full Text] [Related] [New Search]