These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A highly sensitive and selective electrochemical sensor for determination of Cr(VI) in the presence of Cr(III) using modified multi-walled carbon nanotubes/quercetin screen-printed electrode. Author: Sadeghi S, Garmroodi A. Journal: Mater Sci Eng C Mater Biol Appl; 2013 Dec 01; 33(8):4972-7. PubMed ID: 24094212. Abstract: A novel screen-printed carbon electrode modified with quercetin/multi-walled carbon nanotubes was fabricated for determination of Cr(VI) in the presence of excess of Cr(III) without any pretreatment. The method is based on accumulation of the quercetin-Cr(III) complex generated in situ from Cr(VI) at the modified electrode surface in an open circuit followed by differential pulse voltammetry detection. The new method allowed selective determination of Cr(VI) in the presence of Cr(III). The influence of various parameters affecting the adsorptive stripping voltammetry performance was investigated. Under the optimum conditions, the calibration plot was found to be linear in the Cr(VI) concentration range from 1.0 to 200 μmol(-1) with a limit of detection(S/N=3) of 0.3 μmol L(-1). The relative standard deviation (RSD%) of seven replicates of the current measurements for a 50 μmol(-1) of Cr(VI) solution was 3.0%. The developed electrode displayed a very low or no sensitivity to alkali, alkali-earth and transition metal cations and was successfully applied for the determination of Cr(VI) in drinking water samples.[Abstract] [Full Text] [Related] [New Search]