These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Comparisons of choroidal thickness of normal eyes obtained by two different spectral-domain OCT instruments and one swept-source OCT instrument. Author: Matsuo Y, Sakamoto T, Yamashita T, Tomita M, Shirasawa M, Terasaki H. Journal: Invest Ophthalmol Vis Sci; 2013 Nov 19; 54(12):7630-6. PubMed ID: 24168999. Abstract: PURPOSE: We compared the subfoveal choroidal thickness (SFCT) measured on the images obtained by two spectral-domain optical coherence tomographic (SD-OCT) instruments and one swept-source OCT (SS-OCT) instrument. METHODS: A cross-sectional, prospective noninterventional study was done in which SFCT was measured in the images obtained by two SD-OCT instruments; Heidelberg Spectralis-OCT (Spectralis-SD-OCT) and Topcon 3D OCT-1000 Mark II (Topcon-SD-OCT). Images also were obtained with SS-OCT Atlantis DRI OCT-1 (DRI-SS-OCT). After manual segmentation, the measurements were made using the calipers embedded in each instrument. The intrarater, interrater, and intermachine agreements were assessed. RESULTS: We studied 35 subjects. The intrarater correlation coefficient (95% confidence interval) was 0.994 (0.988-0.994) for Spectralis-SD-OCT, 0.996 (0.993-0.998) for Topcon-SD-OCT, and 0.997 (0.991-0.998) for DRI-SS-OCT (P < 0.001). The interrater correlation coefficient was 0.995 (0.991-0.998) for Spectralis-SD-OCT, 0.995 (0.990-0.998) for Topcon-SD-OCT, and 0.996 (0.992-0.998) for DRI-SS-OCT (P < 0.001). The average SFCT was 273.2 μm with Spectralis-SD-OCT, 269.1 μm with the Topcon-SD-OCT, and 280.5 μm with DRI-SS-OCT. The intermachine correlation coefficient was 0.982 (0.964-0.991) for Spectralis-SD-OCT versus Topcon-SD-OCT, 0.907 (0.815-0.953) for Topcon-SD-OCT versus DRI-SS-OCT, and 0.911 (0.832-0.954) for DRI-SS-OCT versus Spectralis-SD-OCT (P < 0.001). The SFCT measured with DRI-SS-OCT was significantly thicker than that with Topcon-SD-OCT, with a mean difference of 11.41 ± 30.27 μm (P = 0.032). CONCLUSIONS: In normal adult eyes, there was good reproducibility and repeatability of SFCT measurements obtained by the SD-OCT and SS-OCT instruments. However, the choroid measured with DRI-SS-OCT was thicker than that measured with both SD-OCT instruments, and, thus, the choroidal thickness should not be compared between the SD-OCT and SS-OCT instruments. (www.umin.ac.jp/ctr number, UMIN000011259.).[Abstract] [Full Text] [Related] [New Search]