These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Tachycardia mediated by an AV universal (DDD) pacemaker triggered by a ventricular depolarization. Author: Pimenta J, Soldá R, Pereira CB. Journal: Pacing Clin Electrophysiol; 1986 Jan; 9(1 Pt 1):105-7. PubMed ID: 2419839. Abstract: A 44-year-old female had Chagas' disease with right bundle branch block, right axis deviation, and ventricular ectopic beats. An invasive evaluation showed a prolonged HV interval of 100 ms. A Medtronic Versatrax 7000 was implanted, after which a pacemaker-mediated tachycardia was observed. The esophageal lead ruled out the possibility of an atrial triggered tachycardia. The diagnosis was pacemaker-mediated tachycardia due to abnormal sensing of the terminal forces of the ventricular activation because of malposition of the atrial lead in the right ventricular out-flow tract.[Abstract] [Full Text] [Related] [New Search]