These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Ion-pair ultra-high performance liquid chromatographic analysis of monoamines: peak-splitting at high flow rates.
    Author: Van Schoors J, Brouwer HJ, Maes K, Michotte Y, Van Eeckhaut A.
    Journal: J Chromatogr A; 2013 Dec 20; 1321():73-9. PubMed ID: 24238712.
    Abstract:
    The use of ion-pair ultra-high performance liquid chromatography (UHPLC) coupled with electrochemical detection (ECD) is of great interest for the fast and sensitive determination of the monoamine neurotransmitters dopamine, noradrenaline and serotonin in microdialysis samples. However, when applying high flow rates in ion-pair UHPLC, other peaks than the initial compound peaks appear on the chromatogram. This peak-splitting phenomenon is caused by disturbed ion-pair retention mechanisms. The influence of several chromatographic parameters is investigated. Peak-splitting is delayed to higher flow rates when increasing the concentration of ion-pair reagent or buffering agent in the mobile phase, when decreasing the percentage of organic modifier in the mobile phase, when applying a stationary phase with a smaller amount of packing material or when increasing the separation temperature. One or a combination of these conditions can be applied to analyze the monoamine neurotransmitters using ion-pair UHPLC-ECD at high flow rates.
    [Abstract] [Full Text] [Related] [New Search]