These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Aptamer-based biosensor for detection of phenylalanine at physiological pH. Author: Omidinia E, Shadjou N, Hasanzadeh M. Journal: Appl Biochem Biotechnol; 2014 Feb; 172(4):2070-80. PubMed ID: 24326680. Abstract: A simple, sensitive aptamer-based biosensor for the detection of phenylalanine is developed using the electrochemical transduction method. For this proposed aptasensor, a 5-thiol-terminated aptamer is covalently attached onto a gold electrode. At the first time, the electrode was evaluated as an electrochemical aptasensor for determination of phenylalanine in aqueous solutions. This sensor was tested in a Tris-HCl buffer with physiological pH = 7.4 by cyclic voltammetry and differential pulse voltammetry. The detection limit and sensitivity of the modified electrode toward phenylalanine were estimated to be 1 nM (S/N = 3) and 0.367 μA nM(-1), respectively. The linear range of the signal was observed between 1 and 10 nM of phenylalanine with 0.9914 correlation factor. The herein-described approach is expected to promote the exploitation of aptamer-based biosensors for protein assays in biochemical and biomedical studies.[Abstract] [Full Text] [Related] [New Search]