  • Title: Effective DQE (eDQE) and dose to optimize radiographic technical parameters: a survey of pediatric chest X-ray examinations in Korea.
    Author: Park HS, Kim YS, Park OS, Kim ST, Jeon CW, Kim HJ.
    Journal: Radiol Med; 2014 Apr; 119(4):231-9. PubMed ID: 24337757.
    Abstract:
    OBJECTIVE: The purpose of this study was to investigate the effect of various technical parameters for dose optimization in pediatric chest radiological examinations by evaluating effective dose and effective detective quantum efficiency (eDQE). MATERIALS AND METHODS: For tube voltages ranging from 40 to 90 kV in 10 kV increments at the focus-to-detector distance (FDD) of 100, 110, 120, 150, 180 cm, the eDQE was evaluated at same effective dose. RESULTS: The eDQE was considerably higher without the use of the grid on equivalent effective dose. This indicates that the reduction of scatter radiation did not compensate for the loss of absorbed effective photons in the grid. The eDQE increased with increasing FDD because of the greater effective modulation transfer function (eMTF) with lower focal spot blurring. However, most of the major hospitals in Korea employed a short FDD of 100 cm with the grid. The entrance surface air kerma values for the hospitals of this survey exceeded the Korean reference level of 100 μGy. CONCLUSIONS: The different reference levels might be appropriate for the same examination conducted on children of different ages. Also, it is necessary to refine the technical parameters to perform pediatric chest examinations.
    [Abstract] [Full Text] [Related] [New Search]