These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Dynamically phase-cycled radial balanced SSFP imaging for efficient banding removal. Author: Benkert T, Ehses P, Blaimer M, Jakob PM, Breuer FA. Journal: Magn Reson Med; 2015 Jan; 73(1):182-94. PubMed ID: 24478187. Abstract: PURPOSE: Balanced steady-state free precession (bSSFP) imaging suffers from banding artifacts due to its inherent sensitivity to inhomogeneities in the main magnetic field. These artifacts can be removed by the acquisition of multiple images at different frequency offsets. However, conventional phase-cycling is hindered by a long scan time. The purpose of this work is to present a novel approach for efficient banding removal in bSSFP imaging. THEORY AND METHODS: To this end, the phase-cycle during a single-shot radial acquisition of an image was dynamically changed. Thus, each projection is acquired with a different frequency offset. Using conventional radial gridding, an artifact-free image can be reconstructed out of this dataset. RESULTS: The approach is validated at clinical field strength [3.0 Tesla (T)] as well as at ultrahigh field (9.4T). Robust elimination of banding artifacts was obtained for different imaging regions, including brain imaging at ultrahigh field with an in-plane resolution of 0.25 × 0.25 mm(2). Besides banding artifact-free imaging, the applicability of the proposed technique for fat-water separation is demonstrated. CONCLUSION: Dynamically phase-cycled radial bSSFP has the potential for banding-free bSSFP imaging in a short scan time, in the presence of severe field inhomogeneities and at high resolution.[Abstract] [Full Text] [Related] [New Search]