  • Title: Guideline for determining the mean glandular dose according to DIN 6868-162 and threshold contrast visibility according to the quality assurance guideline for digital mammography systems.
    Author: Sommer A, Schopphoven S, Land I, Blaser D, Sobczak T, Deutsche Referenzzentren für Mammographie Version 2.0.
    Journal: Rofo; 2014 May; 186(5):474-81. PubMed ID: 24557600.
    Abstract:
    As part of the physico-technical quality assurance of the German breast cancer screening program, the threshold contrast visibility and the average glandular dose of every digital mammography system have to fulfill the requirements of the "European guidelines for quality assurance in breast cancer screening and diagnosis" (4th Edition). To accomplish uniform measurements in all federal states of Germany, the physical board of the reference centers developed a special guideline in 2009. Due to recent changes in the guidelines and standards, a second version of the guideline was developed by the reference centers. This guideline describes the determination of the average glandular dose as well as the CDMAM image acquisition and the CDMAM image evaluation. The determination of the threshold contrast visibility can be performed visually or automatically. The determination of the average glandular dose is based on DIN 6868 - 162 and the threshold contrast visibility test is based on the German "Quality Assurance Guideline".