These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Paediatric CT optimisation utilising Catphan® 600 and age-specific anthropomorphic phantoms. Author: Santos J, Batista Mdo C, Foley S, Paulo G, McEntee MF, Rainford L. Journal: Radiat Prot Dosimetry; 2014 Dec; 162(4):586-96. PubMed ID: 24567497. Abstract: The purpose of the study is to perform phantom-based optimisation of paediatric computed tomography (CT) protocols and quantify the impact upon radiation dose and image noise levels. The study involved three Portuguese paediatric centres. Currently employed scanning protocols for head and chest examinations and combinations of exposure parameters were applied to a Catphan(®)600 phantom to review the CT dose impact. Contrast-noise ratio (CNR) was quantified using Radia Diagnostic(®) tool. Imaging parameters, returning similar CNRs (<1) and dose savings were applied to three paediatric anthropomorphic phantoms. OsiriX software based on standard deviation pixel values facilitated image noise analysis. Currently employed protocols and age categorisation varied between centres. Manipulation of exposure parameters facilitated mean dose reductions of 33 and 28 % for paediatric head and chest CT examinations, respectively. The majority of the optimised CT examinations resulted in image noise similar to currently employed protocols. Dose reductions of up to 33 % were achieved with image quality maintained.[Abstract] [Full Text] [Related] [New Search]