These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: MDCT interpretation of the ascending aorta with semiautomated measurement software: improved reproducibility compared with manual techniques.
    Author: Ahmed S, Zimmerman SL, Johnson PT, Lai H, Kawamoto S, Horton KM, Fishman EK.
    Journal: J Cardiovasc Comput Tomogr; 2014; 8(2):108-14. PubMed ID: 24568947.
    Abstract:
    BACKGROUND: Treatment decisions for ascending aortic aneurysms are guided by measurements from CT angiograms. OBJECTIVE: The aim was to evaluate the reproducibility of these measurements by using manual techniques and advanced imaging software. METHODS: Two radiologists measured maximal ascending aorta diameter on CT angiograms in 30 subjects at 4 separate reading sessions-2 with manual techniques and 2 with semiautomated software analysis. Interobserver and intraobserver variabilities were assessed with Bland-Altman plots and Spearman correlation coefficients. RESULTS: Interobserver variability was smaller for the software-assisted method. Limits of agreement for the manual method were -4.2 mm and 9.2 mm compared with -4.0 mm and 4.6 mm for the software-assisted method; coefficients of repeatability were 6.8 mm and 4.3 mm. Intraobserver variability was inconsistent between readers. Strong correlation was found between observers who used both methods (R(2) = 0.8078-0.9881; P < .05 for all). CONCLUSION: The use of an advanced imaging software for measurement of ascending aortic aneurysm reduces interobserver variability.
    [Abstract] [Full Text] [Related] [New Search]