These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Evaluation of an automated microplate technique in the Galileo system for ABO and Rh(D) blood grouping.
    Author: Xu W, Wan F, Lou Y, Jin J, Mao W.
    Journal: Clin Lab; 2014; 60(2):241-4. PubMed ID: 24660536.
    Abstract:
    BACKGROUND: A number of automated devices for pretransfusion testing have recently become available. This study evaluated the Immucor Galileo System, a fully automated device based on the microplate hemagglutination technique for ABO/Rh (D) determinations. METHODS: Routine ABO/Rh typing tests were performed on 13,045 samples using the Immucor automated instruments. Manual tube method was used to resolve ABO forward and reverse grouping discrepancies. D-negative test results were investigated and confirmed manually by the indirect antiglobulin test (IAT). RESULTS: The system rejected 70 tests for sample inadequacy. 87 samples were read as "No-type-determined" due to forward and reverse grouping discrepancies. 25 tests gave these results because of sample hemolysis. After further tests, we found 34 tests were caused by weakened RBC antibodies, 5 tests were attributable to weak A and/or B antigens, 4 tests were due to mixed-field reactions, and 8 tests had high titer cold agglutinin with blood qualifications which react only at temperatures below 34 degrees C. In the remaining 11 cases, irregular RBC antibodies were identified in 9 samples (seven anti-M and two anti-P) and two subgroups were identified in 2 samples (one A1 and one A2) by a reference laboratory. As for D typing, 2 weak D+ samples missed by automated systems gave negative results, but weak-positive reactions were observed in the IAT. CONCLUSIONS: The Immucor Galileo System is reliable and suited for ABO and D blood groups, some reasons may cause a discrepancy in ABO/D typing using a fully automated system. It is suggested that standardization of sample collection may improve the performance of the fully automated system.
    [Abstract] [Full Text] [Related] [New Search]