Title: Determination of vanadium in groundwater samples with an improved kinetic spectrophotometric method. Author: Bağda E. Journal: Environ Technol; 2014; 35(9-12):1165-74. PubMed ID: 24701912. Abstract: A kinetic catalytic method has been developed for the determination of vanadium based on its catalytic effect on the redox reaction of azorubin S and bromate in the presence of a sulphuric and nitric acid mixture. The reaction is monitored spectrophotometrically by measuring the decrease in absorbance of the reaction mixture at 515 nm. The fixed-time method was used for 0.5-5 min. Optimization of the reaction conditions regarding concentrations of acids, dye, oxidant, masking agent, etc. was investigated. The rate of decrease in absorbance of azorubin S was proportional to the concentration of vanadium in the range of 2.0-1.05 x 10(3) ng mL(-1). 3Sb/m was 0.0129 ng mL(-1) and 10 Sb/m was 0.0432 ng mL(-1). The catalytic method based on the oxidation reaction of azorubin S and bromate shows a good selectivity for vanadium over a wide variety of interference cations and anions. The proposed method was successfully applied to the determination of vanadium in groundwater samples and spiked-water samples.[Abstract] [Full Text] [Related] [New Search]