These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Rapid identification of falsely elevated serum cardiac troponin I values in a stat laboratory. Author: Li Y, Wang X, Xu L, Wen X. Journal: Lab Med; 2014; 45(1):82-5. PubMed ID: 24719992. Abstract: OBJECTIVES: To establish a rapid method for the stat laboratory to identify falsely elevated serum cardiac troponin I (cTnl) values caused by endogenous interferences. METHODS: Serum specimens with elevated cTnI were recentrifuged and cTnI was measured again. If the difference between first and second cTnI value was less than +/- 30%, the specimen was diluted 1:2, 1:5, and 1:10, and the cTnI was measured in the diluted specimens. If the cTnI concentrations did not change linearly with dilution, the specimen was referred to an outside lab for re-analysis of cTnI by a different immunochemical method. RESULTS: Troponin I was elevated in 132 serum specimens. In 18 specimens, we attributed the elevation to the presence of fibrin or microparticles in the serum. Serial dilutions of 3 of the 114 remaining specimens produced nonlinear results. Overall, falsely elevated cTnI results were identified in 1.01% of all measurements. CONCLUSIONS: Our method, which uses combined measurement of cTnI, creatinine kinase MB isoenzyme (CK-MB), and myoglobin, together with incubation, re-centrifugation, and serial dilution, is a practical method to identify falsely elevated serum cTnI values.[Abstract] [Full Text] [Related] [New Search]