  • Title: Moissanite anvil cell design for Giga-Pascal nuclear magnetic resonance.
    Author: Meier T, Herzig T, Haase J.
    Journal: Rev Sci Instrum; 2014 Apr; 85(4):043903. PubMed ID: 24784622.
    Abstract:
    A new design of a non-magnetic high-pressure anvil cell for nuclear magnetic resonance (NMR) experiments at Giga-Pascal pressures is presented, which uses a micro-coil inside the pressurized region for high-sensitivity NMR. The comparably small cell has a length of 22 mm and a diameter of 18 mm, so it can be used with most NMR magnets. The performance of the cell is demonstrated with external-force vs. internal-pressure experiments, and the cell is shown to perform well at pressures up to 23.5 GPa using 800 μm 6H-SiC large cone Boehler-type anvils. (1)H, (23)Na, (27)Al, (69)Ga, and (71)Ga NMR test measurements are presented, which show a resolution of better than 4.5 ppm, and an almost maximum possible signal-to-noise ratio.