These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: the Targon femoral neck hip screw versus cannulated screws for internal fixation of intracapsular fractures of the hip: a randomised controlled trial. Author: Griffin XL, Parsons N, Achten J, Costa ML. Journal: Bone Joint J; 2014 May; 96-B(5):652-7. PubMed ID: 24788501. Abstract: We compared a new fixation system, the Targon Femoral Neck (TFN) hip screw, with the current standard treatment of cannulated screw fixation. This was a single-centre, participant-blinded, randomised controlled trial. Patients aged 65 years and over with either a displaced or undisplaced intracapsular fracture of the hip were eligible. The primary outcome was the risk of revision surgery within one year of fixation. A total of 174 participants were included in the trial. The absolute reduction in risk of revision was of 4.7% (95% CI 14.2 to 22.5) in favour of the TFN hip screw (chi-squared test, p = 0.741), which was less than the pre-specified level of minimum clinically important difference. There were no significant differences in any of the secondary outcome measures. We found no evidence of a clinical difference in the risk of revision surgery between the TFN hip screw and cannulated screw fixation for patients with an intracapsular fracture of the hip.[Abstract] [Full Text] [Related] [New Search]