These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Standardization of methods for measuring plasminogen activator inhibitor activity in human plasma. Author: Chandler WL, Loo SC, Nguyen SV, Schmer G, Stratton JR. Journal: Clin Chem; 1989 May; 35(5):787-93. PubMed ID: 2498016. Abstract: We have standardized the measurement of plasminogen activator inhibitor type 1 (PAI-1) activity in plasma. One-chain tissue-type plasminogen activator (t-PA; EC 3.4.21.31; final activity, 5 int. units/mL) was incubated with plasma (final dilutions 1:4 to 1:40) in phosphate buffer (pH 7.4, ionic strength = 0.15) for 15 min at 37 degrees C, followed by acidification and measurement of residual t-PA activity by an amidolytic method. The PAI-1 activity assay was 98% specific for PAI-1 activity in samples from both pregnancy and nonpregnancy, and varied linearly with added plasma volume when the percent inhibition of t-PA was between 8% and 50%. For the standardized method, analytical recovery was 93 +/- 5%, the detection limit was 1.6 arbitrary units per milliliter (1 arb. unit of PAI-1 activity = inhibition of 1 int. unit of t-PA activity), and total imprecision was 10.2 (SD 0.7) arb. units/mL (CV = 7%, n = 20). The average PAI-1 activity in 10 healthy individuals drawn between 0800 and 1000 hours was 23.9 +/- 15.4 arb. units/mL. Compared with the standardized assay, two of three previously described assays underestimated PAI-1 activity in plasma by 77% and 85%, respectively.[Abstract] [Full Text] [Related] [New Search]