These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Water vapor calibration errors in some capnometers: respiratory conventions misunderstood by manufacturers? Author: Severinghaus JW. Journal: Anesthesiology; 1989 Jun; 70(6):996-8. PubMed ID: 2499226. Abstract: Incorrect calibration has been included in several recently introduced CO2 analyzers. They display a value of "PCO2" internally calculated as FCO2 X Pb rather than FCO2(Pb - 47) where Pb is barometric pressure. This conceptual error appears to have been introduced because new sampling catheter material has become available that effectively removes water vapor before samples reach the sample cell. This seems to have led some manufacturers to assume, incorrectly, that the 47 mmHg factor used to compute PCO2 in patients would no longer be needed. Users can test whether this error is present in an instrument by testing the effect of wet versus dry gases, and make appropriate corrections if the errors are present. Manufacturers should promptly correct this error in all instruments sold previously.[Abstract] [Full Text] [Related] [New Search]