  • Title: The marginal fit of selective laser melting-fabricated metal crowns: an in vitro study.
    Author: Xu D, Xiang N, Wei B.
    Journal: J Prosthet Dent; 2014 Dec; 112(6):1437-40. PubMed ID: 25001421.
    Abstract:
    STATEMENT OF PROBLEM: The selective laser melting technique is attracting interest in prosthetic dentistry. The marginal fit is a key criterion for fixed restorations. PURPOSE: The purpose of the study was to evaluate the marginal fit of cast cobalt-chromium alloy crowns versus the fit of selective laser melting-fabricated crowns. MATERIAL AND METHODS: The marginal gap widths of 36 single crowns (18 selective laser melting-fabricated cobalt-chromium metal crowns and 18 cobalt-chromium cast crowns) were determined with a silicone replica technique. Each crown specimen was cut into 4 sections, and the marginal gap width of each cross section was evaluated by stereomicroscopy (× 100). The Student t test was used to evaluate whether significant differences occurred in the marginal gap widths between the selective laser melting-fabricated and cast cobalt-chromium metal crowns (α=.05). RESULTS: The mean marginal gap width of the cast crowns (170.19 μm) was significantly wider than that of the selective laser melting-fabricated crowns (102.86 μm). CONCLUSIONS: Selective laser melting-fabricate cobalt-chromium dental crowns found improved marginal gap widths compared with traditional cast crowns.