These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Transcutaneous CO2 monitoring in adults with sleep-related breathing disorders.
    Author: Gislason T, Sandhagen B, Boman G.
    Journal: Ups J Med Sci; 1989; 94(2):171-81. PubMed ID: 2503914.
    Abstract:
    The accuracy of transcutaneous CO2 monitoring (PtcCO2) was studied in 22 subjects suspected of having sleep-related breathing disorders, by comparison with arterial CO2 measurements (PaCO2). At rest 40 simultaneous sets of PaCO2 and PtcCO2 were obtained. The mean PaCO2 (+/-SD) was 5.3 +/- 0.9 kPa and PtcCO2 was 5.7 +/- 1.0 kPa (r = 0.79). The ventilatory response to CO2 was evaluated by a CO2 rebreathing method, and simultaneous measurements of PaCO2, PtcCO2 and end-tidal PCO2 (PETCO2) were made every min. Both PaCO2 and PETCO2 increased more during the first min of CO2 rebreathing than PtcCO2 (p less than 0.001). Between 1 to 5 min after the start of rebreathing there were no significant differences between the three methods. During sleep there was an increase in PtcCO2 (by 0.1-0.3 kPa) with each apneic event, the magnitude of the increase depending on the length and distribution of these events. With repeated long apneas there was a cumulative increase in PtcCO2, especially during REM sleep. Continuous PtcCO2 monitoring proved useful in monitoring and diagnosing sleep-related breathing disorders.