These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Electrochemical oligonucleotide-based biosensor for the determination of lead ion.
    Author: Jarczewska M, Kierzkowska E, Ziółkowski R, Górski L, Malinowska E.
    Journal: Bioelectrochemistry; 2015 Feb; 101():35-41. PubMed ID: 25042900.
    Abstract:
    The possibility of utilization of gold electrodes modified with short guanine-rich ssDNA probes for determination of Pb(2+) was examined. Interaction between guanine residues and lead ion followed by formation of G-quadruplex structures was confirmed by electrochemical impedance spectroscopy investigations. An external cationic redox label, methylene blue, was employed in voltammetric measurements for analytical signal generation. It was shown that due to the G-quadruplex formation, the oligonucleotides in the recognition layer fold, which enhances the electron transfer between methylene blue and the electrode surface. The MB current signal rises proportionally to the lead ion concentration in the range from 0.05 to 1μmol/L. The developed biosensor demonstrated high selectivity towards Pb(2+) ion, with only minor response towards interfering metal cations. The calculated limit of detection was of 34.7nmol/L. The utilization of the biosensor for Pb(2+) determination in real samples of water was also tested.
    [Abstract] [Full Text] [Related] [New Search]