These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Electromagnetic thermotherapy for deep organ ablation by using a needle array under a synchronized-coil system.
    Author: Huang SC, Kang JW, Tsai HW, Shan YS, Lin XZ, Lee GB.
    Journal: IEEE Trans Biomed Eng; 2014 Nov; 61(11):2733-9. PubMed ID: 25055378.
    Abstract:
    Thermal ablation by using electromagnetic thermotherapy (EMT) has been a promising cancer modality in recent years. It has relatively few side effects and has therefore been extensively investigated for a variety of medical applications in internal medicine and surgery. The EMT system applies a high-frequency alternating electromagnetic field to heat up the needles which are inserted into the target tumor to cause tumor ablation. In this study, a new synchronized-coil EMT system was demonstrated, which was equipped with two synchronized coils and magnetic field generators to provide a long-range, penetrated electromagnetic field to effectively heat up the needles. The heating effect of the needles at the center of the two coils was first explored. The newly designed two-section needle array combined with the synchronized-coil EMT system was thus demonstrated in the in vitro and in vivo animal experiments. Experimental data showed that the developed system is promising for minimally invasive surgery since it might provide superior performance for thermotherapy in cancer treatment.
    [Abstract] [Full Text] [Related] [New Search]