These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of the enantiomeric and diastereomeric impurities of RS-glycopyrrolate by capillary electrophoresis using sulfated-β-cyclodextrin as chiral selectors. Author: Zuo L, Zhao Y, Ji F, Zhao M, Jiang Z, Sun T, Guo X. Journal: Electrophoresis; 2014 Dec; 35(23):3339-44. PubMed ID: 25104517. Abstract: A practical chiral CE method, using sulfated-β-CD as chiral selector, was developed for the enantioseparation of glycopyrrolate containing two chiral centers. Several parameters affecting the separation were studied, including the nature and concentration of the chiral selectors, BGE pH, buffer type and concentration, separation voltage, and temperature. The separation was carried out in an uncoated fused-silica capillary of (effective length 40 cm) × 50 μm id with a separation voltage of 20 kV using 30 mM sodium phosphate buffer (pH 7.0, adjusted with 1 M sodium hydroxide) containing 2.0% w/v sulfated-β-CD at 25°C. Finally, the method for determining the enantiomeric impurities of RS-glycopyrrolate was proposed. The method was further validated with respect to its specificity, linearity range, accuracy and precision, LODs, and quantification in the expected range of occurrence for the isomeric impurities (0.1%).[Abstract] [Full Text] [Related] [New Search]